A

CONTENTS OF THE PACKAGING

1 mattress, rolled in a transport bag
1 integrated compressor in rolled-up mattress
1 electric power cable which has 2 ways of being attached to the bed.
1 user manual
B

MEANINGS OF THE COMPRESSOR PICTOGRAMS

C

DESCRIPTION

The mattress consists of 18 air cells 12 cm in height placed on a 5 cm foam base.
The cells are inflated alternately (one out of two) except for the 3 head cells, which remain static.
The whole is protected by an integral, removable, waterproof and breathing cushion, incorporating a pillow.
D

HOW IT WORKS

The alternating pressures prevent prolonged vascular compression which may lead to histohypoxia.
The low pressure static mode enables you to handle patients who need to be immobilised (fractures, neurological trauma, etc.) in
order to reduce to a minimum local trauma, helping the patient to rest or carrying out transfer preparations before setting up a
static mattress (...).
The care mode helps with the ergonomics of some care procedures.
The inflation level can be adjusted automatically. You use the pressure adjustment to adapt the mattress to the patient's
morphology and position. Pressure should be calculated when the patient is lying down and calm.
Caution, read the user manual and/or instructions for use
Keep dry
Class II appliance (double-insulated)
Caution electrical and electronic equipment - only to be disposed of in a
special waste collection
Type BF electrical appliance
Compliant with the essential criteria in European Directive 93/42/CEE
applicable to medical devices
Alternating current