AMD RCS-100 Посібник користувача - Сторінка 9
Переглянути онлайн або завантажити pdf Посібник користувача для Медичне обладнання AMD RCS-100. AMD RCS-100 12 сторінок. Multipurpose camera & scope system
11 Cleaning and disinfection
The device is a precision photoelectronic instrument that
should be handled with care.
Please note the following cleaning instructions:
• Turn off the device before cleaning it.
• Disinfect the control unit and charging adapter with the
soft cloth with little alcohol (70% ethyl alcohol). Wait for the
cleaning liquid to evaporate before turning the power on
and connecting the charging adapter and USB cable to the
control unit.
• It is recommended to clean the optical lens with a cleaning
cloth or lens cleaning tissue, such as THORLABS Inc.
(www.thorlabs.com) lens cleaning tissue.
Clean the position of dermatoscope lens which have been in
contact with patient before each use:
• Disinfect dermatoscope lens head with the soft cloth with
alcohol (70 % ethyl alcohol). Wait for the cleaning liquid to
evaporate before attaching with handset.
Replace the specula before each use. If a replacement for
the specula is needed, please contact the manufacturer or
retailer.
Note:
The device is not intended to be sterilized.
12 Troubleshooting
• Failed lens recognition: When the main interface on the
screen does not match the connected lens, disconnect
the lens to confi rm whether the lens connection pins and
surfaces is abnormal or not. After confi rming, connect the
lens again.
• Illumination does not work: Please test the other lenses
and confi rm whether the illumination can be controlled.
• The handset can not turn on: Confi rm that the battery is
charged.
• Battery operating time is too short: Check the battery is in
good condition.
13 Information about the device
RCS-100 is a portable and multifunctional electronic imaging
diagnostic system. Which consists of a handset (5.0 inch, 720
p, multi-touch display, 3.6 V 2.600 mAh rechargeable battery,
1000 photo storage capacity), three kinds of replaceable len-
ses (Otoscope, Dermatoscope, General).
14 Technical data
Size & weight
Size:225mm*135mm*45mm
Weight:292.0 g
LCD
5.0 inch touch (110.7 mm* 62.3 mm) 1280x720
Focus
automatic/manual
Image/video format
picture: JPEG, video: MP4
Battery
3.7V 2600mAH 18650 li-battery
Video 3.5 hours (fully charged battery at 25
ambient temperature)
Adapt
input 100~240V 50~60Hz 0.3A
output DC 5V/2A
CMOS pixels
8M
USB
OTG and type-C USB
RAM
2G LPDDR3
Rom
16G
Extend Memory(OPT)
16G wifi SD card
Otoscope
Size & weight
Size:73.5mm*40.0mm*40.0mm
F/#
2.9
Max object distance
15mm, At max object distance FOV diameter: 15mm
Object height
10mm (Ø4 specula)
Depth of field scope
10mm
Lighting source
natural light LED
21
Dermatoscope
Size & weight
F/#
Polarization:
Object distance
Field of view:
Object distance
Magnification
Field of view:
Photography source
Magnification
LED color temperature:
Photography source
LED color temperature:
General
Size & weight
General
Size & weight
F/#
F/#
Field angle
Object distance
Field angle
Lighting source
Object distance
Operation condition:
• For indoor use only
• Ambient temperature: 10°C to +40°C
• Relative humidity: 15 % to 95 %, rel. non-condensing
• Atmospheric pressure: 700 hPa ~ 1060 hPa
• Storage condition:
• Ambient temperature: 0°C to +45°C
• Relative humidity: 15 % to 95 %, rel. non-condensing
Note:
It is recommended to remove the battery if the device is
stored over 2 weeks.
15 Environment
• Follow the local governing ordinances and recycling plans
regarding disposal or recycling of device components.Es-
pecially when disposing of the lithium ion battery, circuit
board, plastic parts that contain brominated fl ame retar-
dant, LCD, or power cord, be sure to follow the local gover-
ning ordinances.
• When disposing of packing materials, sort them by mate-
rial and follow local ordinances and recycling regulations.
• Inappropriate disposal may contaminate the environment.
• When disposing of specula, or contact plate, follow the
disposal procedures for medical waste such as needles,
infusion tubes, metal instruments for surgery as specifi ed
by your medical facility to avoid infection outside the facility
and environmental pollution.
16 Standards
Electrical safety IEC 60601-1 2005 (EN 60601-1 2006)
EMC and regulatory compliance IEC60601-1-2:2014
ELECTROMAGNETIC COMPATIBILITY
ACCOMPANYING DOCUMENTS ACCORDING TO
IEC 60601-1-2, 2014, Ed. 4.0
The instrument satisfi es the requirements for electromagne-
tic compatibility. Please note that under the infl uence of un-
favorable fi eld strengths, e.g. during the operation of wirel-
ess telephones or radiological instruments, adverse effects
on function cannot be excluded.
The electromagnetic compatibility of this device has been
verifi ed by test according to the IEC60601-1-2:2014 require-
ments.
Weight:96g
Size:62.1mm*62.1mm*36.0mm
Weight:108.5g
2.2
0mm
Dermis / Epidermis
30 mm Ø
0mm
2.5x
30 mm Ø
natural light LED
2.5x
4000K
natural light LED
4000K
Size:60.5mm*60.5mm*19.0mm
Weight:66.4g
2.0
Size:60.5mm*60.5mm*19.0mm
Weight:66.4g
78°
2.0
30mm ~ 4m
78°
natural light LED
30mm ~ 4m