Setting Man down alarm function on/off. User presses the power
A ON/ AOFF
button
(prohibiting Man down alarm).
E
User presses the power button
3. SAN-20 Calibration
This device features one method for calibration: Ambient Air (20.9% Oxygen). This should be
performed in fresh outdoor air. While the device is on, press and hold the temperature unit
button for 5-10 seconds, the monitor will enter the configuration menu. The temperature unit
button can be used to cycle through the configuration and calibration options. When the monitor
displays "AIR" press and hold the power button to start the calibration. The Monitor will display
"IP" during the calibration.
4. Alarm Threshold
There are high and low alarm thresholds in SAN-20. Both high and low alarm has twelve
thresholds: 15.0%, 15.5%, 16.0%, 16.5%, 17.0%, 17.5%, 18.0%, 18.5%, 19.0%, 19.5%, 23.0%,
and 23.5%. Obviously, high alarm threshold should not less than low threshold. They can be
same alarm level.
The SAN-20 is equipped with audible, visual and vibration alarms to alert users when the
ambient oxygen concentration exceeds either of the two factory preset alarm levels:
Danger Low Oxygen Alarm: LED will flash and audible alarm will sound 3x / sec.
Warning High Oxygen Alarm: LED will flash and audible alarm will sound 2x / sec.
To alert users in noise environment, an inner vibrator will vibrate at 1Hz in both Oxygen Alarms.
Normal atmosphere contains between 20.8%and 21% oxygen. OSHA defines any atmosphere
that contains less than 19.5% oxygen as "oxygen deficient" and any atmosphere that contains
more than 23.5 percent oxygen as "oxygen enriched."
ATTENTION! - If a Warning or Danger Alarm is triggered while using the
instrument as a safety monitor, leave the area and seek fresh air immediately.
Remaining on site under such circumstances can cause serious impairment or
5. Automatic Atmospheric Pressure Compensation
The O2 measurement is affected by atmospheric pressure or altitude changing. When users are
at high altitude, compensation should be made to assure maximum monitor accuracy.
This device has automatic atmospheric pressure compensation for O2 concentrations by means
of a built-in barometric pressure sensor.
6. Man down alarm
Copyright © CO2Meter, Inc. All Rights Reserved
to switch "A ON" (allowing Man down alarm) or "AOFF"
even lead to death.
to exit the menu status.