PREPARING TO RUN AN OPTIONAL LIQUID QUALITY CONTROL TEST
1. Gather the Following Items
• CoaguChek XS Pro Meter
• Container of test strips
• Test Strip Code Chip
• CoaguChek XS Pro PT Control
Solutions
• Diluent Dropper (2)
• Quality Control Code Chip
RUNNING AN OPTIONAL LIQUID QUALITY CONTROL TEST
1. Remove a Test Strip
Take a test strip out of the container.
Close the container tightly.
You have 10 minutes to use a test strip
once you remove it from the container.
6. Collect Control
Using the dropper, gently squeeze
and hold the bulb. Insert the tip of the
dropper into the control solution. Slowly
release the bulb to draw the control up
to the bulb. Avoid creating bubbles in the
control.
CCXS_Pro_GSG_05914647001_03_USA.indd 2
2. Match Code
3. Reconstitute Control Solution
The code number on the control solution
Open a bottle of lyophilized control.
and the Quality Control Code Chip must
Cut off the flared tip of the dropper at
match.
the end of the stem.
Each box of controls comes with a
Hold the diluent dropper vertically by
matching code chip. Control code chips
the stem, not the bulb, to avoid loss of
are identified with a "C" in the number.
diluent.
Do not hold the pipette close to
your face.
2. Insert Test Strip
3. Select Control Test
Main Menu
03/16/2010
Slide the test strip into the test strip
Press Control Test on the main menu
guide in the direction of the arrows until
screen.
it stops.
An hourglass symbol appears.
The meter powers on, and the main
The QC TEST menu appears.
menu appears.
7. Apply the Control
8. Record Result
Apply a drop of control to the target area.
Record your result as instructed by
your facility.
Hold the control drop to the test strip until
you hear a beep. The flashing dropper
Note: If during testing the meter displays
symbol will disappear. An hourglass will
an error message, refer to the Error
appear, indicating the test is in progress.
Messages section of the user manual
for an explanation and steps on how to
Return any unused control to the bottle for
proceed.
additional tests. Reconstituted controls are
Run a level 2 control test with a new test
good for 30 minutes.
strip following the same steps.
The result appears in less than one
minute.
Transfer the entire contents of the
dropper to the bottle. Do not touch the
dropper to the dried material or to the
sides of the bottle.
Do not discard the dropper. You will
need it to apply a drop of control to the
test strip later in the test.
4. Insert QC Code Chip
09:15 am
QC Test
*
New
Patient Test
QC 248
Control Test
QC 379
Review Results
QC 462
Setup
QC 332
03/16/2010
Press NEW on the QC Menu.
A code chip symbol with a flashing
arrow appears on the screen.
Remove the Test Strip Code Chip and
insert the Quality Control Code Chip with
the number facing up until it snaps into
place.
The meter stores the code chip
information in memory for future use.
9. Remove Control Code Chip
Remove the Quality Control Code Chip
and store it with the remaining controls.
Replace the Test Strip Code Chip.
Roche Diagnostics recommends that you
keep a code chip in the slot to keep the
electrical contacts free of dirt.
Close the bottle.
Gently swirl the bottle using a circular
motion to dissolve the dried material.
Do not shake or turn the bottle on its
side or upside down. Avoid creating
bubbles.
Let the bottle sit undisturbed for at least
a minute to ensure that the dry material
dissolves completely.
5. Select the Level
09:15 am
QC Test
09:15 am
QC:
999
Code: 184
Par:
PT
Op:
Operator 1
Level 1
Level 2
03/16/2010
Select Level 1 for running a level one
control.
An hourglass symbol appears.
A test strip symbol appears with a
flashing dropper symbol. You have
180 seconds to apply the quality control
solution.
10. Clean Up
Place the used strip and control bottles in
an approved biohazard container.
Power the meter off.
If the meter is dirty, wipe it clean.
COAGUCHEK is a trademark of Roche.
©2007-2016 Roche Diagnostics.
All rights reserved.
Roche Diagnostics GmbH
Sandhofer Strasse 116
68305 Mannheim, Germany
Made in Germany
Manufactured for and distributed in the USA by:
Roche Diagnostics
9115 Hague Road
Indianapolis, IN 46250-0457
Rx Only
www.coaguchek-usa.com
www.usdiagnostics.roche.com
06.05.2016 11:56:13