dealt with by advising the technical parameters or a stilling tube.
2.1.3 hints for solid surface conditions
For fine-grained solids, the sensor must be aligned with the product
surface.
2.1.4 hints for in-tank effects
Stirrers or agitators can cause a vortex. Mount the transmitter off-center
of any vortex to maximize the return echo.
In non-linear tanks with rounded or conical bottoms, mount the
transmitter off-center. If needed, a perforated reflector plate can be
installed on the tank bottom directly under the transmitter center line
to ensure a satisfactory return echo.
Avoid mounting the transmitter directly above pumps because the
transmitter will detect the pump casing as the liquid falls away.
When install to the cold area, should choose the lengthen sensor of the
level meter ,make the sensor extend into the container, shun frost
and icing .