Supervisory Mode and Calibration Instructions
for IPI Industrial Pressure Indicator
Using the Supervisory Mode
Push the CONFIG button to access the user-settable functions on the
indicator. Each time the CONFIG button is pressed; the display
advances to the next function. Press CONFIG repeatedly until "FUnC
LOCK" is displayed.
Press ENTER when "FUnC LOCK" is displayed, 0
on the indicator. The password to enter supervisory mode is 101, set
using the ▲ and ▼ keys. Holding a key continuously will cause the
display to advance more quickly for faster setting. The password is
factory set and cannot be changed.
1. Your IPI is shipped from the factory with all setting access
"unlocked" or available to be changed.
2. In supervisory mode each of the parameters can be locked or
unlocked using the ▲ and ▼ keys. Select LOC (lock) for those
parameters you do not want to be accessible, and UnLOC (unlock)
for those can be accessed.
3. In order, the functions that can be unlocked, locked or accessed
are:
● Zero function (enable/disable)
● Set pressure units (enable/disable)
● Auto shutdown adjustment (enable/disable)
● Damping settings (enable/disable)
● Sample rate setting (enable/disable)
● Tare setting (enable/disable)
● Custom engineering units (set scale factor)
4. Use the CONFIG key to scroll through the above choices, and the
▲ and ▼ keys to lock and unlock features. Press CONFIG to
continue scrolling through the parameters, pressing ENTER at any
point saves your settings and returns the indicator to normal
operation.
PWRD
will be displayed