CohuHD Costar 3212-2000 Посібник із швидкого старту - Сторінка 4

Переглянути онлайн або завантажити pdf Посібник із швидкого старту для Вимірювальні прилади CohuHD Costar 3212-2000. CohuHD Costar 3212-2000 17 сторінок. Thermometric system

CohuHD Costar 3212-2000 Посібник із швидкого старту

Outdoor Mounted configuration

If using the camera in an outdoor setting, provide an enclosed area (tent) to ensure accuracy as
illustrated below.

Physical Installation Considerations

1. For accurate temperature measurement (+ 0.3ºC / 0.5ºF Degrees / Medical grade accuracy) install
the camera in areas without strong air flow. Place the camera away from
airflow or high ambient temperature. Do not install the camera opposite an exterior door, air
conditioner, or in a position allowing the sun to shine directly onto the camera.
To ensure accuracy, install the camera indoors or in a stable temperature-controlled environment. If
you install the thermographic system in an open-air environment, temperature readings may differ
from actual temperatures, requiring a software adjustment to compensate.
2.
If installed outdoors, provide a tent or an enclosed environment to ensure accuracy.
3.
It is necessary to perform temperature measurements within a standardized distance, otherwise
temperature readings will deviate. The best horizontal distance from the camera to a person's face is
13 feet (4m), with an acceptable range of 10–16 feet (3–4.8m).
CohuHD 3212-2000 Thermometric System Quick Start Guide
objects that produce
3