COKEVA 11940C Інструкція з експлуатації - Сторінка 18

Переглянути онлайн або завантажити pdf Інструкція з експлуатації для Вимірювальні прилади COKEVA 11940C. COKEVA 11940C 19 сторінок. Close field probe

Cokeva 11940C  Operation Note
 
1.    Use a 7/32" wrench to remove the old SMA Adapter (see Figure 10). Turn the adapter counterclockwise 
 as viewed looking at the rear of the probe. 
 
2.    The SMA Rotary Joint has one SMA connector located at each end. One connector rotates with respect to the 
body of the Rotary Joint. The other connector is fixed to the body. Cover two or three end threads of the fixed end 
connector with a small amount of Loctite 222®. 
 
3.    Screw the rotating end of the Rotary Joint clockwise onto the probe's SMA Connector (see Figure 11). 
 
4.    Using a 9/32" torque wrench, tighten the Rotary Joint to 24 inch‐pounds. 
 
5.    When attaching a cable to the Rotary Joint, hold the Rotary Joint by hand to keep the connector from turning. 
 
 
 
13