3 3/4 DIGIT AUTO DIGITAL MULTIMETER
OPERATION MANUAL
SAFETY RULES
λ
This meter is designed and tested in accordance with IEC publication 1010, pollution
degree II and installation category (overvoltage category) III 600V.
λ
This meter has been tested according to the following EC Directives
♦
89/336/EEC (EMC of Nov., 1992, Electromagnetic Compatibility)
♦
73/23/EEC (Product safety law of June 11, 1979, Low Voltage Directive of
February 19, 1973)
λ
This meter is designed to be indoor use at temperature 41°F to 104°F (5°C to 40°C)
and altitude up to 2,000m.
λ
To ensure that the meter is used safely, follow all safety and operating instructions in
this operation manual. If the meter is not used as described in this operation manual,
the safety features of this meter might be impaired.
λ
Do not use the meter if the meter or test leads look damaged, or if you suspect that the
meter is not operating properly.
λ
When using the probes, keep your fingers behind the finger guards on the probes.
λ
Disconnect the live test lead before disconnecting the common test lead.
λ
Make sure power is off before cutting, unsoldering, or breaking the circuit. Small
amount of current can be dangerous.
λ
Do not apply more than 600Vdc or 600Vac rms between a terminal and earth ground.
λ
To avoid electrical shock, use CAUTION when working above 60Vdc or 25Vac rms.
Such voltages pose a shock hazard.
λ
Never make measurements with the battery cover or bottom case off.
λ
To avoid electrical shock or damage to the meter, do not exceed the input limits.
INTERNATION SYMBOLS
Important information
!
see manual
AC
DC
FEATURES
λ
One rotary switch operation.
λ
Auto polarity with engineering units annunciator
λ
Dual polarity A-D converter.
λ
Auto voltage ranges selection: DCV: 1mV - 600V, ACV: 1mV - 600V
λ
Auto ampere ranges selection: DCA: 40mA - 10A, ACA: 40mA - 10A
λ
Auto resistance ranges selection: 0.1Ω - 40MΩ
λ
Diode & continuity test
3.1 General Specifications
Display
: 3 3/4 digit LCD, 3999 counts with decimal point and
engineering units annunciator display
Polarity
: Automatic, (-) negative polarity indication.
Zero adjustment
: Automatic.
Measuring Method
: Dual slop integration A to D converter system.
Over range indication
: Only the "OL" is displayed.
Power
: Two, standard 1.5 volt (AAA) battery, UM4 or equivalent.
Dimension
: 70 (H) x 144 (W) x 40 (D) mm.
Weight
: Approx. 280gm. (including batteries and packaging)
3.2 Electrical Specifications
Accuracies are ± (% of reading + number of least significant digits) at 23°C ± 5°C, <75% RH.
DC Voltage
Range
Resolution
4V
1mV
40V
10mV
600V
1V
AC Voltage
Range
Resolution
4V
1mV
40V
10mV
400V - 600V
1V
DC Current
Range
Resolution
10A
10mA
AC Current
Range
Resolution
10A
10mA
Resistance
Range
Resolution
400Ω
100mΩ
4kΩ
1Ω
40kΩ
10Ω
400kΩ
100Ω
4MΩ
1kΩ
40MΩ
10kΩ
Diode Test
Range
0~3.999V
Continuity
Range
Test
400Ω
PANEL DESCRIPTIONS
1) 3 3/4 digit LCD display
2) Continuity test
3) Resistance range
4) AC voltage range
5) DC voltage range
6) Power off
7) Rotary switch
8) Data hold / power resume button
9) Case
10) Diode range
11) DC current range
12) AC current range
AM-34
Diode
Continuity
Ground
Double insulation
Accuracy
Input
Overload
Impedance
Protection:
±(0.8%+1)
10MΩ
600V DC/AC rms
Accuracy
Frequency
Input Impedance:
(Hz)
10MΩ
Overload
±(1.2%+3)
40-400
Protection:
±(1.5%+5)
40-100
600V rms
Accuracy
Input resistor
Overload
±(1.8%+5)
Protection:
0.01Ω
Accuracy
Input resistor
10A/600V fuse
±(3.0%+5)
0.01Ω
Accuracy
Measurement
Response time:
Voltage/Current
5 sec. max. on
VDDA 1.2mA or
40MΩ range
less
3 sec. on all other
0.7V(TYP)
range
±(1.0%+2)
0.7mA or less
Temperature range
for 40MΩ from
0.47V(TYP)
18°C to 30°C
0.7mA or less
±(1.5%+2)
Resolution
Open short or Normal Range
1mV
Resolution
Buzzer sound generation resistance
100mΩ
_50Ω
OPERATION WARNING
1) When measuring voltage ensure that the instrument is not connected or switched to a
current or resistance or temperature or to the diode check/continuity range. Always ensure
that the correct terminals are used for the type of measurement to be made.
2) Use extreme care when measuring voltage above 50V, especially from sources where
high energy exists.
3) Avoid making connections to "live" circuits whenever possible.
4) When making current measurements ensure that the circuit is not "live" before opening it
in order to connect the test leads.
5) Before making resistance measurements or continuity / diode test, ensure that the circuit
under test is de-energised.
6) Always ensure that the correct function and range is selected. If in doubt about the correct
range, start with the highest and work downwards.
7) Extreme care should be taken when using the instrument to conjunction with a current
transformer connected to the terminals. High voltage may be produced at the terminals if
an open circuit occurs.
8) Ensure that the test leads and prods are in good condition with no damage to the
insulation.
9) Take care not to exceed the over-load limits as given in the specifications.
10) Fuse for replacement must be of the correct type and rating.
5.1 DC and AC Voltage measurement
1) Connect the black test lead to the "COM" socket and red test lead to the "
2) Set the selector switch to DC V or AC V position and connect the test prods across the
source or load under measurement.
5.2 DC and AC Current measurement
1) Connect the black test lead to the "COM" socket and red test lead to the "10A" socket for
measurement up to 10A.
2) Set the selector switch to the DC current or AC current position.
3) Connect the prods in series with the current source to be measured.
5.3 Resistance measurement
1) Connect the black test lead to the "COM" socket and red test lead to the "
2) Set the selector switch to resistance Ω position.
3) Connect the prods across the circuit to be tested.
!
CAUTION: Ensure that the circuit to be tested is "dead".
Max. input over-load : 600V rms < 10sec.
5.4 Diode test
1) Connect the black test lead to the "COM" socket and red test lead to the "
2) Set the selector switch to
position.
3) Connect the black and red test prods to the cathode (-) and anode (+) ends of the diode to
be tested respectively.
4) Read the forward voltage drop (junction) value from the display. If reverse connected the
prods to diode, display shows over-load.
5.5 Continuity test
1) Connect the black test lead to the "COM" socket and red test lead to the "
2) Set the selector switch to
position.
3) Connect the prods across the circuit to be tested, if the resistance less than approx. 50Ω,
buzzer will be activated.
5.6 Data hold
A push switch at the middle of rotary switch is used to hold data during measurement.
Pressing the key, reading is held and "H" appears on display. Resume active reading by
pressing switch gently once.
5.7 Auto power off
The meter will turn off automatically to extend the battery life after approx. 15 minutes without
operation. Resume power by pressing data hold push switch gently once.
6. MAINTENANCE
!
CAUTION
BEFORE ATTEMPTING BATTERY AND FUSE REMOVAL OR REPLACEMENT,
DISCONNECT TEST LEADS FROM ANY ENERGISED CIRCUITS TO AVOID SHOCK
HAZARD.
6.1 Fitting and replacing the battery and fuse
1) Ensure that the instrument is not connected to any external circuit, set the selector switch
to OFF position and remove the test leads from the terminals.
2) Remove the screw of the back case.
3) Replace the spent battery or fuse with the same type and rating.
4) Reinstate the back case, tighten and securing screw.
6.2 Cleaning
Periodically wipe the case with a soft damp cloth and mild household cleanser. Do not use
abrasives or solvents. Ensure that no water gets inside the equipment to prevent possible
shorts and damage.
FOR TECHNICAL ASSISTANCE,
PLEASE CONTACT:
AMPROBE
Tel : 305-423-7500
Fax : 305-423-7554
Technical Support : 800-327-5060