waste water tank and make sure the exhaust hose is in the above of the
waste water tank, but not in the water. If you are using " TR" Series,
please empty the water inside the waste water tank or only leave a
little bit, because the water dried will be drained to the tank.
5. Add water into the chamber, make sure it is below the water plate, but
not too little, you can observe through the hole of the water plate.
6. Put the sterilizing articles in the baskets then put it on the water plate
of the chamber.
7. Close the door: turn the handle clockwise until the " Locked" light is
on, not blinks, must be stably on. (The machine will not start if the
door is not properly locked)
8. Selecting the sterilizing program: Press the "UP" and "DOWN" button
to choose the sterilizing program. The No.8 button will show the U01
to U04, this is the set programs. If you want to sterilize under those
programs. You just need to press the start for 3 seconds to start. If you
want to create the new programs, you can press "SET/ENT" button to
enter into the setting menu, press the "UP" and "DOWN" to set the
temperature, then press the "NEXT" to enter into time setting. Then
pressure "SET/ENT" again to save the new program, the new
programs will be U05 to U20.