DN:KD-926-SMSY01 V1.0
SPECIFICATIONS
1. Product name: Wireless Blood Pressure Monitor
2. Model: KD-926
3. Classification: Internally powered, Type BF applied part,IP20,No AP or APG, Continuous
operation
4. Machine size: Approx. 145 mm ×90 mm × 51mm
5. Cuff circumference: 8-21/32" to 16-17/32"(22cm-42cm)
6. Weight: Approx. 234g (8 1/4oz.) (exclude cuff)
7. Measuring method: Oscillometric method, automatic inflation and measurement
8. Memory volume: 2×60 times with time and date stamp
9. Power source: batteries: 4x1.5V
10. Measurement range:
Cuff pressure: 0-300mmHg
Systolic:
Diastolic:
Pulse rate:
11. Accuracy:
Pressure:
Pulse rate:
12. Environmental temperature for operation: 10° C-40° C(50° F -104° F)
13. Environmental humidity for operation: ≤85%RH
14. Environmental temperature for storage and transport: -20℃~50℃(-4℉~122℉)
15. Environmental humidity for storage and transport: ≤85%RH
16. Environmental pressure: 80kPa-105kPa
17. Battery life: Approx 300 times.
18. Wireless Connection:
Smart Bluetooth
Frequency Band: 2.400~2.4835GHz
19. All components belonging to the pressure measuring system, including: Pump,Valve, LCD,
Cuff, Sensor
Note: These specifications are subject to change without notice.
NOTICE
1. Read all of the information in the operation guide and any other literature in the box before
operating the unit.
2. Stay still, calm and rest for 5 minutes before blood pressure measurement.
3. The cuff should be placed at the same level as your heart.
4. During measurement, neither speak nor move your body and arm.
5. Measuring on same arm for each measurement.
6. Please always relax at least 1 or 1.5 minutes between measurements to allow the blood
circulation in your arm to recover. Prolonged over-inflation (cuff pressure exceed 300
mmHg or maintained above15 mmHg for longer than 3 minutes) of the bladder may cause
ecchymoma of your arm.
7. Consult your physician if you have any doubt about below cases:
SIZE AA, DC 6V 600mA
60-260mmHg
40-199mmHg
40-180 beats/minute
±3mmHg
±5%