Droplet Genteel Інструкція із застосування - Сторінка 2

Переглянути онлайн або завантажити pdf Інструкція із застосування для Лабораторне обладнання Droplet Genteel. Droplet Genteel 2 сторінки. Lancing device

Droplet Genteel Інструкція із застосування
Installing the Contact Tip onto the Nozzle
1
Place the Contact Tip on a clean, hard, flat surface,
like a tabletop.
2
There are two sides to the Contact Tip—the smooth side that
goes against your skin and the side with the grooves that goes
onto the Nozzle. Make sure the side with the groove is facing up.
Use a slight back and forth (rotating) motion to gently push the
Nozzle flat into the Contact Tip. The lip on the Nozzle goes into
the groove on the Contact Tip.
3
The Contact Tip may also be installed by stretching it over
the Nozzle.
Testing from the fingertip
4
Select area to be lanced and wipe with an alcohol swab
and/or tissue.
5
Cock Droplet Genteel by pushing in the
Push Cap, taking care not to hold the
Activation Button while priming (you will
hear a click if Droplet Genteel is cocked
correctly).
DO NOT HOLD ONTO THE ACTIVATION BUTTON
WHILE COCKING DROPLET GENTEEL, AS THIS
WILL DISENGAGE THE LOCKING MECHANISM
AND DROPLET GENTEEL WILL NOT COCK.
6
Remove the Nozzle by twisting
counterclockwise, then pull it off.
7
Insert a new, sterile lancet and
carefully twist off its protective cap.
NOTE: Make sure the lancet is inserted
correctly into the lancing device. It is
okay for the lancet to turn slightly
during protective cap removal.
120 Day Satisfaction Guarantee
When you use the Droplet Genteel lancing device with Droplet 30G Personal Lancets and
in accordance with the instructions for use, we offer a 120 day satisfaction guarantee.
Please visit www.mydropletgenteel.tips within 30 days of purchase to register your device.
If you are unsatisfied with your Droplet Genteel lancingdevice for any reason, please call us
at 1-877-660-1900 or email us at [email protected] to receive a one-time replacement
device or refund within 120 days of purchase.
Please do not attempt to return product without prior authorization. Any Droplet Genteel
lancing device returned without prior authorization will be refused.
8
Replace the Nozzle by lining up
the body's pins with the Nozzle
slots and rotate clockwise.
9
Place Droplet Genteel, gently, but firmly, against
the fleshy side portion of your middle, ring, or
pointer finger. Look to be sure there are no
spaces between the Contact Tip and your skin.
10
Press the Activation Button and continue to hold firmly
on the test site. Count to 10 or until you see a blood
drop form when you look inside the Nozzle window.
11
Release the Activation Button.
12
Pull Droplet Genteel away
from your skin
Removing the Lancet
13
Cock Droplet Genteel by pushing in the
Push Cap at the back of the device.
14
Twist and take off the Nozzle.
15
Take out the lancet and place it
into the sharps container.
Alternate site testing
Alternate site testing means using a part of your body other than
the fingertips to obtain blood for blood glucose testing. Always
check with your doctor to see if and when alternate site testing
is right for you.
If alternate site testing is right for you, choose a penetration site
on a soft, fleshy area of your arm, or other recommended site,
away from bone, and free of visible veins and hair. Obtain a
sample following Steps 4-12, the same way as shown in
"Testing from the fingertip".
Manufactured for:
HTL-STREFA S.A.
ul. Adamówek 7
95-035 Ozorków, Poland
T: +48 42 270 00 10
F: +48 42 270 00 20
www.htl-strefa.com
Made in the United States
730/9150/09/2020
20201012_Droplet Genteel Lancing Device_Website IFU
Cleaning Droplet Genteel Lancing Device
After every use
We recommend using a disinfectant on Droplet Genteel to help
prevent the transfer of bloodborne viruses or infectious agents.
We recommend wiping the lancing device, including the inside
and outside of the Nozzle and Contact Tip, with 70% isopropyl
alcohol. After wiping, allow Droplet Genteel to air dry completely.
1 Surface cleaning:
Before every use, wipe the outside of the device with a clean
damp cloth.
IMPORTANT: If the Push Shaft rises past its halfway point –
while holding the Activation Button, this may be a vacuum leak
caused by dirt or other contaminants inside Droplet Genteel.
If you experience this vacuum loss symptom – perform (2).
If symptom persists – perform (3). If cleaning doesn't resolve
vacuum leak, call customer service 1-877-660-1900.
2 Quick clean (inside):
2.1 Remove Piston Assembly: De-prime
Droplet Genteel by pressing Activation
Button. Twist Piston Assembly base
counter-clockwise until you can pull it out
of the Vacuum Tube.
2.2 O-ring Wipe: Wipe the O-ring on the
Piston Assembly with your clean finger to
remove surface dirt or other contaminants.
3 Deep clean (inside):
3.1 Remove Piston Assembly: Repeat step 2.1
3.2 Flush Vacuum Tube: Run fast moving, warm
water through the Vacuum Tube and out the
Nozzle while repeatedly pressing the
Activation Button.
3.3 Flush Piston Assembly: While running the
Piston Assembly under fast-moving, warm
water repeat 2.2.
3.4 Dry: Let Droplet Genteel air dry thoroughly.
Do not reassemble Droplet Genteel with moisture in the
Vacuum Tube.
3.5 Reassemble: Put Piston Assembly back inside the
Vacuum Tube and twist base clockwise until it locks into
place securely. Verify device functions as intended.
Distributed by:
If you need troubleshooting assistance or more
help with your Droplet Genteel lancing device,
HTL-STREFA, Inc.
3005 Chastain Meadows Pkwy,
PLEASE DO NOT WAIT
Suite 300
Marietta, GA 30066, USA
T: +1-770-528-0410
TO CONTACT US.
F: +1-770-528-0411
www.mydroplet.com
WE ARE HERE TO HELP!
1-877-660-1900
2.1
2.2
3.2
3.3