Remove caps, remove any liquid adhering to the chimneys of the
cartridge and remove any bubbles on the surface of the reagent
Page down to locate the position(s) on the CC carousel where
When prompted by the software, open the reagent carousel door
UniCel DxC, Customer, Daily Start Up Job Aid
Rev. F (September 2013)

Load CC Reagents

Prepare reagent(s) if required
Select Rgts/Cal from the menu bar
reagent cartridge(s) will be placed (pages 3 to 8)
Select all Positions to load
reagent cartridges
Select Load F1
Place cartridge in the load position in front of the blinking bar
code reader beam
Enter Reagent name (Chem)
Using the picture on the right as
a reference, enter the following:
Lot number (numbers only)
Serial number
Expiration date (MM/YY)
Select the Tab key on the
keyboard to save the entries
Verify the reagent information is displayed on the monitor
Move the cartridge onto the reagent carousel
Close the reagent carousel door
Additional
reagents
selected?
No
Procedure complete
OR
Yes