Install the new TDF69 (continued).
d. Attach the wire terminals to the white
connector.
As shown in the picture to the right,
hold the white connector so the slotted
side is on the left and facing you. Insert
the wire terminals with the spring
facing away from you and the barb
facing towards you. Insert the black
wire in the top, the white wire in the
middle and the red wire at the bottom.
Lightly tug on each wire to ensure that
the barb is working and the wires do not
pull out.
e. Plug the white connector back into the spot that
is second from the bottom.
f. Feed the vacuum sensor tube through the 1/4"
cable holder.
Note: This may be difficult and require you to
first remove the cable holder.
g. Tighten the nut on the cable holder. If you
removed the cable holder then reattach it using
a nut driver. Be sure to hold the cable holder
when tightening to ensure that it does not rotate.
h. Connect the vacuum sensor 1/4" silicone tube to
the barbed T-fitting that is connected to the
EtOH78 supply valve.
ANKOM Technology 2052 O'Neil Road Email: [email protected] Phone: 315-986-8090 Fax: 315-986-8091
Vacuum Sensor Installation
TDF
White connector
Spring
Barb
White Connector
Nut
Cable holder
Vacuum
sensor
1/4" tube
T-fitting
013
Service Procedure
Revised: 07/16/2014
pg. 3 of 4