Drucker 642VES Посібник оператора - Сторінка 3
Переглянути онлайн або завантажити pdf Посібник оператора для Лабораторне обладнання Drucker 642VES. Drucker 642VES 12 сторінок. Laboratory centrifuge
Features:
•
Swing–out horizontal rotor design, incorporating a unique test tube holder that produces horizontally
separated samples while requiring no additional parts
•
Cool–Flow air flow design that prevents overheating of samples
•
Lid safety switch that prevents the centrifuge from operating unless the lid is closed and latched
•
Removable rotor for easy cleaning
•
Locking lid that allows entry into the centrifuge only after the rotor has completely stopped
•
Clear lid for safe observation of samples and optical calibration of speed
•
Electronically controlled timed operation
•
Push-button operation
•
Indicator Lights
'RUNNING'
'RUN 1' VFD-Plus only
'RUN 2' VFD-Plus only
'LATCHED'
'UNLOCKED'
Specifications 642VFD, 642VFD-Plus, and 642VES:
Speed Range:
(Horizontal)
(Fixed Angle)
RCF Range:
(Horizontal)
(Fixed Angle)
Maximum Capacity:
Overall Dimensions (H x W x D):
Weight:
Centrifuge Motor:
Nominal Acceleration Time:
Protection:
Timer (642 VES):
Timer (642 VFD, 642VFD-Plus):
Power Rating:
Voltage Requirements:
Frequency:
Permitted Environmental Conditions
Ambient Temperature During Operation:
Maximum Relative Air Humidity:
* Maximum sample density is 1.15 grams / mL, (water density = 1.0 grams / mL)
Any use other than those specified by the Manufacturer is explicitly prohibited.
Green – lights when power is applied to motor
Green – lights when power is applied to motor
Green – lights when power is applied to motor
Yellow – lights when lid is closed and latched
Red – lights when the locking system is deactivated
Page 3
1,000 - 3,800 (+/- 100) RPM
1,000 – 3,900(+/- 100) RPM
114 – 2016 (+/- 90) xg
90 – 1860 (+/- 90) xg
60 mL (6 × 10 mL)*
8.75 in. × 11.75 in. × 14 in.
(22 cm x 30 cm x 35 cm)
12 lbs. (5.4 kg)
1/30 HP, PSC motor
20 seconds
4 Amp, resettable circuit breaker (x2)
Electronic, 1 to 99 minutes, preset
to 10 minutes +/- 1%
Electronic, 1 to 30 minutes, preset
to 10 minutes +/- 1%
200 Watts
115/ 230 VAC (+/- 10%)
50/60 Hz
60°F - 90°F (16°C - 32°C)
90%