S
E
UPPLIED
QUIPMENT
One (1) six–place horizontal rotor
Six (6) 75/100mm tube holders
The rotor and accessories are rated for a rotation frequency of 4,000 RPM.
For optional accessories, see page 13 of this manual.
F
EATURES
•
Swing–out horizontal rotor design incorporating a unique test
tube holder that produces horizontally separated samples while
requiring no additional parts
•
Cool–Flow air flow design that prevents overheating of samples
•
Heavy gauge steel construction for safety and durability
•
Lid safety switch that prevents the centrifuge from operating
unless the lid is closed and latched
•
Removable rotor for easy cleaning
•
Brushless permanent split capacitor AC motor
•
Clear lid for safe observation of samples and optical calibration
of speed
G
S
ENERAL
PECIFICATIONS
Nominal Speed (Horizontal)
Nominal RCF (Horizontal)
Maximum capacity (Horizontal)
Overall Dimensions (H x W x D)
Centrifuge Motor:
Maximum Acceleration Time
Protection Breaker
Timer
Current Requirement
Voltage Requirement
Frequency
Weight
Any use other than those specified by the Manufacturer is explicitly prohibited.
Maximum sample density is 1.15 grams / mL, (water density = 1.0 grams / mL)
Page | 3 Drucker Diagnostics +1-814-692-7761 - [email protected]
P/N
7786061
P/N
7713079
3,380 (+/- 100) RPM
1,600 (+/- 90) xg
60 mL (6 x 10 mL)
8.75 in. x 11.75 in. x 14 in.
1/30 HP, p.s.c. motor
10 seconds
4 Amp. re–settable
electronic, 1 to 30 minutes,
+/– 10%
1.5 Amps
115 (+/- 10) Volts
60 Hz
11 lbs