3.3 Disinfection and sterilization of torque wrench and Endo wrench
3.3.1 Torque wrench and Endo wrench can be sterilized with any commonly used non-corrosive neutral disinfectant, or they can be sterilized by
autoclaving (Please refer to 3.2.1).
3.3.2 Do not disinfect or sterilize the torque wrench in the following ways:
a) Put it in solution and cook it;
b) Soak with disinfectant such as iodine, alcohol and glutaraldehyde;
c) Bake it at high temperatures in an oven or in a microwave oven.
Note: if the torque wrench is directly or indirectly damaged as a result of the above three disinfecting and sterilizing methods, Woodpecker will not
be responsible for the repair.
3.4 The tip is made of 3Cr13 and has obtained Biocompatibility Test Report which shows that it will not result in any
unacceptable risk while touching teeth.
3.5 Cleaning of tips, endochuck, torque wrench and endo wrench
The scaling tip, endochuck, torque wrench and endo wrench can be cleaned by ultrasonic cleaner.
3.6 Troubleshooting and notes
3.6.1 Troubleshooting
Fault
The scaling tip doesn't
The power plug is in loose contact.
vibrate and there is no
The foot switch is in loose contact.
water flowing out when
The fuse in the adapter or main unit is broken.
stepping on the foot switch.
The tip is in loose contact.
The scaling tip doesn't
The connect plug between the handpiece and the circuit board
vibrate but there is water
is in loose contact.
flowing out when stepping
Something wrong with the handpiece.
on the switch.
Something wrong with the cable.
Possible cause
Solutions
Make the plug insert to the socket well.
Insert the foot switch to its socket tightly.
Contact our dealers or us.
Screw the tip on the handpiece tightly (See Picture 4).
Contact our dealers or us.
Send the handpiece to our company to repair.
Contact our dealers or us.