5.9.4 Demonstrating communication software

(1)The super end under WIN9X/NT/2000/XP
(2)The debugging assistant on gjwtech.com
(3)Self-arranging communication software

5.9.5 Order Input

6.

Care and Maintenance

To keep the balance operating properly, the housing and platform should be
kept clean and free from foreign material. If necessary, a cloth dampened with
a mild detergent may be used. Keep calibration masses in a safe dry place.
Unplug the AC Adapter when not in use. For long term storage, remove the
batteries.
7.

Error Codes

The following list describes the various errors which may appear on the display
and the suggested remedy:
Err 0 With the platform on, this signal indicates the platform is too light. Make it
heavier properly. If this error still occurs, the sensor or the circuit board may be
destroyed. The balance must be sent to the manufacturer to be serviced.
Err 1 The sample being weighed exceeds the capacity of the balance. If the error
occurs when the sample is within the balance capacity, maybe the balance is
not correctly calibrated.