Anyload S2 Посібник користувача - Сторінка 10
Переглянути онлайн або завантажити pdf Посібник користувача для Вимірювальні прилади Anyload S2. Anyload S2 15 сторінок.
available for decimal point setting.
Cap: Capacity,Specifies the full scale of the indictor, press Enter to save
the capacity after inputting numbers.
Capacity=Full Scale Value/Display Division.
e.g. The full scale value is 5000kg, the decimal point is 1, Division is 1,
so the Capacity=5000/0.1*1=50000.
Zero pos: Calibrate the zero position of the scale.
Calibr: Calibrate the full scale of the scale, there is a instantaneous
value on the right side, when the weighing is stable, there will be a [Stb.]
The calibration of S2 consists of the following steps:
Zero calibration
Enter the Full Scale Value
Capacity calibration
3.1.1 ZERO Calibration
STEP1.Remove all load from the scale platform. If the scale require hooks or
chains (tare weight), place the hooks or chains onto the scale for zero
calibration.
STEP2.Press 4/L.F to calibrate zero, the screen shows "completed!" 5.Calibr
will show 0[Stb.] on the right side.
3.1.2 Enter the full scale weight value
Press 5/AUTO, the 5 digits number will turn to 00000, now the value is editable,
input the full scale value.
3.1.3 Capacity calibration
Place full load onto the scale. When [Stb.] is shown, press Enter to complete
the capacity calibration.
3.2 Calibration Parameter
ZERO pos: Display zero calibration A/D count, not editable.
Parameter: Display capacity calibration A/D count, not editable.
3.3 Password
Change the password by inputting 6 digits, press Enter to save.
3.4 Power saving mode
Press 4/L.F to select this item, if there is a tick in the end, the power saving
mode is enabled.
In power saving mode, the indicator will be shut down automatically if no
Page 10 of 15 AL‐Manual‐110ES‐WL‐S2‐En‐103