reading to settle. Repeat this step to get as close as possible to 2.000
mV.
2.000 mV is 2 Amps actual current flow out of the battery.
12. Write the mV reading into the calibration work sheet. (2 Amp
calibration)
13. Press NEXT to save the calibration data into memory. Wait for the
word "SAVED" to appear then disappear.
14. Press the UP button to select "Amps: 8.5".
15. Press NEXT to start the process. The FAN will start.
16. Wait 30 seconds.
17. Press the UP or DOWN button to obtain a DMM reading of 8.483 to
8.517 mV (8.500 mV ±0.017 mV). Allow a few seconds for the
reading to settle. Repeat this step to get as close as possible to
8.500mV.
8.500 mV is 8.5 Amps actual current flow out of the
battery.
18. Write the mV reading into the calibration work sheet. (8.5 Amp
calibration)
19. Press NEXT to save the calibration data into memory. Wait for the
word "SAVED" to appear then disappear.
20. Press the UP button to select "Amps: 15.0".
21. Press NEXT to start the process. The FAN will start.
22. Wait 30 seconds.
23. Press the UP or DOWN button to obtain a DMM reading of 14.970 to
15.030 mV (15.000 mV ±0.030 mV). Allow a few seconds for the
reading to settle. Repeat this step to get as close as possible to
15.000mV.
15.000 mV is 15 Amps actual current flow out of the
battery.
24. Write the mV reading into the calibration work sheet. (15 Amp
calibration)
25. Press NEXT to save the calibration data into memory. Wait for the
word "SAVED" to appear then disappear.
26. Press UP to select "Amps: 55.0".
27. Press NEXT to start the process. The FAN will start.
28. Wait 30 seconds. This will allow the current sensor to reach operating
temperature.
29. Press UP or DOWN button to obtain a DMM reading of 54.890 to
55.110 mV (55.00 mV ±0.110 mV). Allow a few seconds for the
reading to settle. Repeat this step to get as close as possible to 55.00
mV.
30. Write the mV reading into the calibration work sheet. (55 Amp
calibration)
31. Press NEXT to save the calibration data into memory. Wait for the
word "SAVED" to appear then disappear.
32. Turn OFF the BC-8000 power.
Page 5