7. Refrigerator Installation
DO NOT
position the refrigerator in direct sunlight or next to any heat sources. Make sure the
refrigerator is positioned to provide good ventilation and ease of access for maintenance.
Ventilation is particularly important for the condensers and compressors.
1)
Position the refrigerator on a level surface; wipe all internal and external surfaces with a
5)
Check that the refrigerator compartment temperature display on the front of the cabinet is
operating. If there is insufficient natural light shine a torch on the display for a few seconds.
clean cloth.
2)
Check the DC isolator at the back of the
cabinet is in the '0 position.
3)
Connect the refrigerator to the Dulas PS-ILR-
001 power supply. Check that there is
sufficient slack in the cable to allow access to
the back of the refrigerator.
4)
Plug the Dulas PS-ILR-001 into a suitable
AC outlet and turn it on. Then switch the DC
isolator to the "1" position. Within 1 minute
the compressor will start if there is AC power.
You will be able to hear the compressor
operating and feel the condensers getting
warm.
6)
Ensure that the lid is firmly closed.
7)
Check that all cables are securely supported so
that they will not be pulled loose if the refrigerator
is moved for cleaning.
8)
Roll the refrigerator into its final position and apply
the brakes on the castors to secure it.
Dulas VC225ILR User Manual