10. User Training
The end user of the VC30SDD should be instructed in the simple processes outlined below and
also those in Section 11 Maintenance tasks. It should be stressed that these tasks are essential to
ensure correct and reliable operation of the refrigerator.
Caution: failure to follow these instructions may lead to system failure and vaccine
wastage.
If you have a problem, please contact your appointed engineer.
When turning on for the first time or when the refrigerator has been turned off for more than 3
days:
Connect the solar panels and allow the refrigerator to run for at least 7 days.
Before loading with vaccines, ensure that the temperature display on the front of the
refrigerator reads +4.5
Daily Routine
Keep the lid closed, open only when required and lock after use.
Do not keep food or drink in the refrigerator.
Vaccine Storage:
Only store vaccines in the refrigerator.
Use the baskets provided: Do not store vaccines outside the basket.
Always store vaccine in its original packaging.
Always keep the same type of vaccine together and store your vaccine neatly.
Always use old vaccines before new ones; before use make sure the vaccine has not
expired.
11. Maintenance Tasks
Every morning and afternoon:
Check the temperature
Fill in the daily record sheet
On the first day of every week:
wipe away any moisture build up around cabinet lid
Clean the solar array
On the first day of every month:
Clean the refrigerator, condensers and compressors and drain any water from the
bottom of the cabinet using in built drain
Check the solar array is not shaded between 7am and 5pm
Every 6 months:
Check all mechanical fixings and electrical connections (including the array)
o
C or less by 4pm in the afternoon.
VC30SDD User Manual