3 Service
This section includes periodic maintenance procedures, a list of replaceable parts, a troubleshooting
voltages table, and a troubleshooting guide.
3.1 Periodic Maintenance
The table below outlines a periodic maintenance schedule for the CM2010 IZS module. This schedule
is based on experience under normal operating conditions, and may need to be modified to suit specific
operating conditions.
1
Interval
6 months
Zero Air Scrubber
DFU Filter
1 year
Purge Filter/Orifice
Flow/Leak Check
1
Suggested intervals for normal operation, and actual intervals will vary depending upon application.
This table can be used as a guideline, but the user should develop a maintenance schedule to suit
their specific requirements.
2
Replacement interval determined by ambient pollutant levels.
HB4014-02
EVOLUTION SERIES CM2010 OZONE ANALYSER
Table 3-1 Routine maintenance
Item
Figure 3-1 Maintenance components
Procedure
2
Check/Replace
Replace
Clean/Replace
Perform
9