5.3 Bulb Replacement

The MAGW200 may need to have its bulb replaced during the lifetime of the light. To
remove the bulb please follow these instructions.
Do not use undue force when replacing lamps, as it may shatter causing
injury.
Ensure the MAGW200 is disconnected from mains before replacing
lamps.
-
Slide the rear cover off, being careful not to bend it.
-
Identify which of the 4 bulbs need replacing.
-
Grasping the metal areas on the left and right of the bulb, turn the bulb 90 degrees
towards you. Notice the green bulb holder on either side of the bulb turn to the open
position, exposing the bulb prongs. See images overleaf.
-
Slide the bulb out carefully, and replace with an identical 4W bulb type (DARAY order
code: DRK4213)
-
Rotate the bulb 90 degrees to the locked position.
-
Slide the rear cover on.
Page 9 of 16