Spark Marathon

6.1 Specific Problems

▌ An error message "Cannot establish communication with ..." appears when opening
Clarity Instrument.
Solution:
Check the power cable (Spark Marathon sampler must be switched on),
communication cable and communication settings (the autosampler must be in
Serial mode, i.e. the "Serial mode" inscription must be shown on the Autosampler
display). If it is not, switch it on.
▌ An error message "AS Error" appears during the Clarity operation.
Solution:
The communication has been interrupted. Check the communication cable as it is
most probably disconnected. This message may also occasionally appear after
aborting the ACTIVE Sequence.
▌ Injection volume set in the
Solution:
Either you are using the Flushed Loop option and the injection volume doesn't the
one of the installed injection loop, or you are trying to enter the volume that is
greater than the half of the installed Loop Volume in the Partial Loopfill injection
mode.
▌ The Spark Marathon autosampler does not start with the injection.
Solution:
Check the Freeze input active settings in the I/O configuration of the Spark
Marathon autosampler. When set to HIGH, the injection will not start until the
respective input is closed. For more details on how to check the I/O configuration
see the Spark Marathon manual.
Sequence
window is not accepted.
- 14 -
6 Troubleshooting