Step 5
Fluoroscopically monitor the radiopaque band
on the distal end of the guiding catheter. When
a sufficient length of the guiding catheter is
above the stricture, disconnect the Luer lock
fitting (the gray connection with black arrows)
on the device.
Step 8
To deploy the stent, retract the guiding catheter into the endoscope while maintaining the position
of the pushing catheter. Confirm the position of the stent once the guiding catheter is removed,
then remove the pushing catheter from the accessory channel.
customer Service
eMeA:
EDI – www.cookmedical.com/edi.do
Distributors: +353 61239240, [email protected]
Austria: +43 179567121, [email protected]
Belgium: +32 27001633, [email protected]
Denmark: +45 38487607, [email protected]
Finland: +358 972519996, [email protected]
France: +33 171230269, [email protected]
Germany: +49 6950072804, [email protected]
Hungary: +36 17779199, [email protected]
Ireland: +353 61239252, [email protected]
Italy: +39 0269682853, [email protected]
Netherlands: +31 202013367, [email protected]
Norway: +47 23162968, [email protected]
Spain: +34 912702691, [email protected]
Sweden: +46 858769468, [email protected]
Switzerland – French: +41 448009609, [email protected]
Switzerland – Italian: +41 448009609, [email protected]
Switzerland – German: +41 448009609, [email protected]
United Kingdom: +44 2073654183, [email protected]
M E D I C A L
www.cookmedical.com
Step 6
Maintain position of the guiding catheter and
advance the stent into the desired position
using the pushing catheter. Fluoroscopically
and endoscopically confirm desired
stent placement.
Americas:
EDI – www.cookmedical.com/edi.do
Phone: +1 812.339.2235, 800.457.4500, Fax: 800.554.8335
E-mail: [email protected]
Australia:
Phone: +61 738411188, 1800777222, Fax: +61 738411288, 1800077283
E-mail: [email protected]
Step 7
Unlock the wire guide from the Wire Guide
Locking Device. Using fluoroscopy, retract the
wire guide until it exits the guiding catheter
at the IDE port. For multiple stent placement,
advance the disengaged wire guide to maintain
ductal access and relock the wire guide.
Step 9
The introducer, separated from the stent.
Refer to current instructions for detailed system use.