7. Maintenance & Troubleshooting
No daily maintenance is required. It is intended this equipment should only be serviced by qualified
and authorized technical personnel. It is suggested that professional healthcare and caregivers to check
patients' conditions and settings every two hours to prevent any emergency or misuse situation.
Fault/Problem
Description
No indication that
the pump is on.
Low pressure in
mattress/mattress
is not inflating.
Pump Controls lock
up 'freeze'.
If problem is not resolved, please contact your sales representative for advice.
CAUTION: Please ensure the system is connected to a stable power or in connection with
Uninterruptible Power Supplies.
Troubleshooting
Check Universal Therapy System (UTS) is
connected to the mains power supply.
Check for loose connection on plug and
main power is switched on.
Check if mains socket is faulty.
Check Universal Therapy System (UTS) is
connected to the mains power supply.
Check Universal Therapy System (UTS)
air connections are fitted securely.
Ensure airflow is coming out from pump.
Check CPR valve is set to CLOSE position
and is not leaking.
Check the connector tubes for kinks,
obstructions or damage.
Check air intake from filter is not blocked
by linen/dust.
Turn off and unplug Universal Therapy
System (UTS)
Solution
Check the power is switched on.
Secure plug connection and turn
on power from main.
Try another socket.
Ensure the main power is turned on
and Universal Therapy System
(UTS) is plugging into main.
Ensure connectors are securely
fastened and reconnect pump air
hoses if loose.
Ensure pump is turned on.
Ensure CPR valve is set to CLOSE
position. Replace a new CPR valve if
an air leak is found.
Undo any kinks and obstructions.
Replace with new filter.
Rest Universal Therapy System
(UTS) for a few seconds and plug
the back in to main and turn on the
pump.
8