4.2. STERILISATION

Before each autoclave cycle lubricate the micromotor with a quick squirt of a lubricant for dental handpieces legally marketed and approved in your
country (see paragraph PRODUCTS TO BE USED).
Spray the lubrificant into the central hole in the back side of the micromotor only.
Reconnect the motor onto the instrument tubing and activate it so as to expel the excess lubricant.
Dry the outside of the micromotor with gauze or cotton wool before proceeding on with sterilisation.
Set following parameter according to Type of sterilizer:
Type of Sterilizer (EU market)
Dynamic-air-removal (e.g., prevacuum)
Type of Sterilizer (USA market)
Gravity displacement
Dynamic-air-removal (e.g., prevacuum)
• For sterilization use a wrap legally marketed and approved in your country (see paragraph PRODUCTS TO BE USED).
• Do not use ultrasound cleaners.
• Do not immerse the handpiece in disinfectant or sanitizing solution.
• Never sterilise the handpiece in a dry heat steriliser.
• Do not leave the handpiece in the autoclave, but remove it immediately after the cycle.
• Periodically check the autoclave according to the requirements of the manufacturer. Temperatures higher than the indicated
limit may damage the micromotor.

4.3. PRODUCTS TO BE USED

The Manufacturer recommends to use the following products:
CEFLA s.c. does not guarantee proper device operation if products other than those indicated are used.
EU market
DETERGENT FLUIDS
LUBRICATING OILS
STERILIZATION WRAPS
USA market
DETERGENT FLUIDS
LUBRICATING OILS
STERILIZATION WRAPS
All the products listed above must be stored and used according to the instructions of their Manufacturers.
EN
8
I-MMr • i-MMr L • i-MMs
Exposure time at 121 °C
20 min
Exposure time at 250° F
30 min
• IC100 (ALPRO MEDICAL GMBH) or other CE marked detergent
• Daily Oil PLUS – 500 ml (CEFLA s.c.)
• Use only CE marked sterilization wraps
• Sani ProZyme™ - 3,8l (Crosstex)
• Statcare™ - 500ml (SciCan)
• H200 - (Halyard Health)
Exposure time at 134 °C
Exposure time at 270° F
/
Drying time
4 min
40 min
Drying time
/
40 min
4 min
40 min