STEP 7
STEP 8
NOTE:
Rev B
Figure 4: Lamp connector
REATTACH the lamp chamber cover with the
thumbscrew. Carefully reposition the outer cover
and reattach the four screws.
PLUG THE UNIT IN AND TURN ON the power
switch. Reset the lamp hour counter by pressing and
holding the up arrow button during warm-up. A
message appears on the display asking if you want
to reset the counter; "YES' or "NO" (flashing "NO" is
default). Select response using left and right arrow
keys and press the green center "Enter" button. If
"YES" is selected, a message will appear on the
display asking if you are sure; "YES' or "NO"
(flashing "NO" is default). Select response using left
and right arrow keys and press the green center
"Enter" button. The lamp hour counter is now reset.
The lamp hour reset menu can only be
accessed during the 3 minute lamp warm-up
period. Pressing and holding the up arrow
button in normal operating mode will not initiate
the lamp hour reset sequence.
Lamp power
connector
7