Activox LifeChoice Посібник оператора - Сторінка 7
Переглянути онлайн або завантажити pdf Посібник оператора для Медичне обладнання Activox LifeChoice. Activox LifeChoice 16 сторінок. Lifechoice activox portable oxygen concentrator
When the LifeChoice Activox POC is being powered from the
internal battery, the Battery Check LED labeled INT will be
slowly flashing. Pushing the Battery Check Button will display
the charge level of the internal battery.
If not connected to the AC or DC Adapter/Charger, the internal
battery will slowly discharge over time even when not in use. Always
check the battery level prior to use to ensure adequate charge level.
2. AC Adapter/Charger: An AC Adapter/Charger allows the LifeChoice
Using an AC Adapter/Charger that is not specified by Inova Labs
Inc. for use with the LifeChoice Activox POC may damage the unit
and will void all warranties.
3. DC Adapter/Charger: A DC Adapter/Charger can be connected
from the LifeChoice Activox POC unit to a vehicle's (car, boat, motor
model cars will not provide sufficient current to operate the POC
and charge the battery at the same time. Verify that any DC outlet
used can provide at least 10 amps at 12 VDC. (Car auxiliary power
varies significantly; therefore the rate of charging will vary.)
4. External Battery: To power the LifeChoice Activox POC from an external
battery, connect the external battery to the external battery connector
port. A fully charged external battery will supply power to the LifeChoice
Activox POC for 3 hours. (Reference Battery Life table since actual time
will vary depending upon breath rate and Pulse Setting.)
When the LifeChoice Activox POC is being powered from the
external battery, the Battery Check LED labeled EXT will be slowly
flashing. Pushing the Battery Check Button will display the charge
level of the external battery.
NOTE: After turning on the LifeChoice Activox POC, allow 30 seconds
for the unit to reach stated performance.
10
Activox POC to be powered from a
standard 100-240 VAC, 50/60 Hz
outlet. When the unit is operated from
an AC Adapter/ Charger, power from
the AC outlet powers the unit and
recharges the internal and/or external
battery. (See Battery Charging
section for details.)
home, etc.) standard 12 VDC outlet.
When the system is connected to the
DC outlet, power from the vehicle battery
powers the LifeChoice Activox POC unit
and recharges the internal and/or
external battery. (See Battery Charging
section for details.)
The 12 VDC outlets in some pre-2000
BATTERY CHARGING
To charge the internal and/or external battery, connect the LifeChoice
Activox POC to either an AC or DC Adapter/Charger and a suitable
outlet. If the external battery is being charged, also connect the external
battery to the External Battery Port. Confirm that one of the Battery
Check LEDs, either INT or EXT, is solid green. This indicates that an
AC or DC external power source has been recognized and that battery
charging is in progress.
If an external battery is connected to the LifeChoice Activox POC, it is
important to understand the battery charging sequence. By design,
the internal battery is charged first. That means that the external battery
will not start charging until the internal battery is charged to a 100%
level. Use the INT and EXT Battery Check LEDs to determine which (if
any) batteries are charging.
To check the battery charge level while charging is in progress, press
the Battery Check Button. The charge level will be displayed as indicated
in the Battery Charge Indicator section (See User Controls). When a
battery is fully charged, the corresponding Battery Check LED (INT or
EXT) will be off.
It takes a minimum of four hours for a fully discharged internal battery
to reach a 100% charge level. If both internal and external batteries
are fully discharged, it can take up to seven hours to achieve a 100%
charge level in both batteries. It is recommended to recharge the
batteries, even if only partially depleted, as often as possible. The
batteries cannot be overcharged, and it is recommended to leave the
LifeChoice Activox POC plugged in via the AC or DC Adapter/Charger
as often as possible.
In the event the power supply is interrupted to the LifeChoice Activox
POC when the AC or DC charger is being used, a two-second audible
alarm will sound to inform you that power has been lost. Check that
the charger has been properly connected to the power supply.
The rechargeable batteries are maintenance-free and can only be
exchanged at the factory.
NOTE: If the LifeChoice Activox POC is powered off and either the AC
or DC Adapter/Chargers are connected to the LifeChoice Activox POC
after charging is complete, the internal fans may continue to operate
for an additional period of time until the temperature inside the unit is
sufficiently cool.
REPRESSURIZATION
If the unit has not been used for a period of one or more days, it will
automatically repressurize itself. During this time, you may hear the
compressors running for a few seconds.
11