Communication Configuration
1. Your computer must be directly connected to the Em50G to
configure its communication.
2. To turn the cellular network of the Em50G on or off, return
to "Configure" and choose "GSM". The cellular network is active
when "upload data to ECH2OData.com" box is checked and is off
when the box is not checked. The Em50G is uploading data to the
Decagon Data Service ("on") by default.
3. Choose up to six upload times. Default upload times are:
6-7a.m., 10-11 a.m., 1-2 p.m., 4-5 p.m. To conserve battery power,
choose fewer upload times.
4. Click "Apply" to save communication settings.
5. Telemetry Test: Check the quality of the cellular communication
by choosing "Connection" , then "Devices tools", then "test". The
test takes approximately 30 seconds so please be patient until the
"test" button is no longer gray.
Installing the Em50G
1. Perform a communication test to confirm that there is good
cellular coverage at your site. If cellular coverage is low, try moving
a few meters away and trying the communication test again.
2. Attach the Em50G in an upright position to a firm post or PVC
pipe, using the five zip ties or similar fasteners.
Test Buttons
The Em50G has a test button and two indicator LEDs above the
batteries for basic functionality testing.
• When you push the "TEST" button on the
OK
logger, both the "ERROR" (red) and "OK" (green)
lights will slowly blink on and off indicating that
ERROR
the testing is in progress.
• When the Em50G is sending data to Decagon's
server, the "OK" (green) light will blink rapidly.
OK
• At the end of the testing, if everything is
working properly, the "OK" (green) light will
ERROR
remain lit for approximately 20 seconds.
• If you have pressed the "TEST" button and the
OK
bottom red light remains lit, a communication
error is occurring. Perform a telemetry test to
ERROR
attempt to diagnose the problem.
Downloading Data
Downloading Data For the First Time
The Em50G sends all of your data to the Decagon server, where
the data are stored. Please download the most up to date version
of DataTrac 3 from the Decagon website (www.decagon.com)
BEFORE trying to download data from our server.
]