The centrifuge must only be opened by trained personnel competent in its use.
•
• Never use the centrifuge unless the rotor is properly mounted and secure.
• Never open, or attempt to open the lid while the centrifuge rotor is still turning.
• Only use original parts for the centrifuge.
• Never operate the centrifuge if components are missing or damaged.
• If the rotor shows visible signs of wear its use must be discontinued and
replacement fitted.
• The rotor must not be overloaded, the maximum allowed density of samples at
full operating speed is 1.2g/cm3.
• Always endeavour to make sure that opposite tubes are of the same weight to
avoid rotor imbalance. If the tubes are filled with the same material they must,
therefore, be filled with the same amount.
• No changes should be made to the mechanical or electrical components without
prior
consultation with and the written permission of Centurion Scientific.
• The centrifuge has not been manufactured from inert materials or to be explosion
proof.
Ensure that it is not operated within an environment, or with materials that makes
these a
requirement.
• During operation a "safe" zone of 30cm must be maintained around the
centrifuge. This zone should be clear of personnel and hazardous materials at all
times during the run.
• Substances of a radioactive, flammable or explosive nature must not be
centrifuged.
• Substances prone to react briskly with each other must not be centrifuged at the
same
time.
• Toxins and pathogenic micro-organisms must not be centrifuged unless the
proper precautions for their handling have been taken. These may include, but are
not limited to, biological seals.
Should toxins or pathogens enter the centrifuge or its parts the proper procedures
for disinfecting the centrifuge should be carried out.
• Strongly corrosive substances that may cause damage to or impair the
mechanical strength of the rotor may only be centrifuged inside protective vessels.
Should any doubt exist regarding the substance or the suitability of a particular
vessel for use with it, consult the Safety Inspector.