DANGER
5 Storage
6

Maintenance

8
INCORRECT MEASURED VALUES AFTER EXCCEEDING OF
MEASURING RANGE
If a CO concentration above the maximum value of the
measuring range was measured, the sensor requires a
recovery time before it can deliver precise measured values
again.
If the maximum value of the measuring range was exceeded,
immediately leave the area where the measured CO
concentration occurred.
If the maximum value of the measuring range was exceeded,
perform a function test by taking a measurement at a point with
a known, safe CO concentration and verifying that the known
CO concentration is correctly displayed.
Do not perform further measurements unless you have verified
that the sensor delivers correct measured values.
Failure to follow these instructions will result in death or serious
injury.
Store the device in a dry location away from solvents.
The measuring instrument does not contain any parts that can be
serviced or repaired by the customer.
Prior to performing a measurement, perform a visual inspection of
the CAPBs for visible damage. Do not use damaged measuring
instruments.
Clean the measuring instrument after each use. Use a dry,
lint-free cloth for cleaning.
Use a slightly wetted cloth to remove pollution that cannot be
removed with a dry cloth.
Do not use cleaning agents for cleaning.
The CAPBs measuring instrument must be serviced by the
manufacturer or an authorised service point at regular intervals. The
service intervals depend, among other things, on the legal
requirements and regulations.
®
CAPBs