DISPOSABLE PATIENT CIRCUIT
Attach High or Low Flow Vapor Transfer
Cartridge to Disposable Water Path.
Attach patient delivery tube to
Disposable Water Path.
Open door and install Disposable Water Path into
docking station.
Hang sterile water bag.
Wipe water spike with alcohol pad and insert into
sterile water bag.
Allow a minimum of 200ml of water to fill into the
Disposable Patient Circuit.
Volumax VOS is ready for start up.
START UP AND ADJUSTING PARAMETERS
Install oxygen sensor.
Install gass inlet filters on back with filter
bowls vertical (glass side down).
Attach air & O
hoses. Plug in power cord.
2
Rotate the blue Setting Control Knob to
illuminate display.
Press in Setting Control Knob to select the parameter
and rotate to adjust the value.
Press and release (do not hold) the Run/Standby button
once to start. A GREEN light indicates RUN mode
(AMBER light indicates STANDBY mode, No Flow).
Green light will stop flashing once temperature
is reached.
CONNECT TO PATIENT
The flashing green LED becomes steady when the
set variables are reached.
Place the cannula on the delivery tube.
After cannula is attached, it is ready to be placed
on patient.
2601 Stanley Gault Pkwy, Suite 101
800-923-8413
502-244-6345
Louisville, KY 40223 USA
FAX: 502-244-0369
BATTERY BACK-UP MODE
To activate the internal backup battery mode,
the Run/Standby LED must be green before
disconnecting the power cord.
The unit will enter into battery mode and will maintain
flow and oxygen percentage for at least 15 minutes
The battery icon will flash.
SHUT DOWN MODE
Press the Run/Standby button. Unit will enter
Standby mode.
Clamp the water inlet tube and disconnect it by
pulling out the spike.
Open the door, remove the Disposable Patient Circuit
(includes delivery tube & cartridge) by sliding it upwards
out of the docking station
Discard all disposables according
to hospital guidelines.
Disconnect unit from AC power.
www.avantevet.com