Allied Healthcare Schuco S130 Посібник із швидкого старту - Сторінка 6
Переглянути онлайн або завантажити pdf Посібник із швидкого старту для Медичне обладнання Allied Healthcare Schuco S130. Allied Healthcare Schuco S130 12 сторінок.
READ AND SAVE THESE INSTRUCTIONS
OPERATION:
WARNING: PROPER USE OF THIS PRODUCT IS NECESSARY TO AVOID PHYSICAL HARM. IMPROPER USE CAN
个
CAUSE PERSONAL INJURY AND/OR TISSUE DAMAGE. THIS PRODUCT SHOULD NOT BE USED UNLESS A
PHYSICIAN OR OTHER QUALIFIED HEALTH CARE PROFESSIONAL HAS INSTRUCTED THE OPERATOR.
10.
The ON/OFF switch should be in the OFF
position.
Plug the line cord into a grounded electrical outlet, making sure that it is the same voltage as indicated on the unit nameplate.
Be sure the length of the tubing from the Disposable Canister Unit (DCU)/glass bottle to the filter is dry.
Turn the ON/OFF switch to the ON position.
The control knob will adjust the vacuum. As the knob is turned counterclockwise, the vacuum will decrease. As the knob is turned
clockwise, the vacuum will increase. Inches of vacuum will register on the gauge. Use only the vacuum setting directed by the Physician or
Oualified Health Care professional. Check degree of vacuum by pinching off patient tube.
The level of vacuum in mm/inches of Hg will register on the vacuum on the gauge.
The pump is now ready for patient use.
The DCU is a throw away item and after complete use or full, turn off the unit, disconnect the tubing, seal the open ports, and dispose of the
DCU properly.
The glass bottle should be emptied after complete use or when full. To empty the bottle, turn the unit off, disconnect the tubing, and remove
the cap and float assembly.
Follow the cleaning procedures listed below. Once these procedures have been completed, see Step 1 under Set-Up before next patient
use.
CLEANING PROCEDURE TO BE USED AFTER EVERY USE:
1.
750-cc Glass Bottle and Lid:
It is recommended that the bottle, lid assembly and tubing be cleaned after every use. To ensure the longest life for these parts the
following procedures must be followed:
Clean the 750-cc autoclavable bottle, lid assembly, gasket, seal washer, float guide and float thoroughly with a mild soap. Rinse
thoroughly with clean water and dry thoroughly.
Place parts in autoclave as follows ensuring the parts are not touching:
e
Bottle with open end down
|
e
Float and float guide upright with large ends down
e
Gasket and seal washer can be placed in a gusset pouch prior to autoclaving
Autoclave for 3 to 5 minute cycle at 270° F (132° C).
In lieu of autoclave sterilization, the above parts can be soaked in an activated dialdehyde solution.
Vacuum Pump:
|
人 个
| WARNING: DO NOT SUBMERGE VACUUM PUMP IN WATER!!!
|
|
Make sure the unit is unplugged; wipe the exterior surface with a mild disinfectant or a solution of 70% Isopropyl Alcohol solution. Wipe all
surfaces with a soft cloth or disposable wipe moistened in the cleaning solution.