CPS TEMP-SEEKER TM360 Посібник з експлуатації - Сторінка 8

Переглянути онлайн або завантажити pdf Посібник з експлуатації для Вимірювальні прилади CPS TEMP-SEEKER TM360. CPS TEMP-SEEKER TM360 13 сторінок. Precision thermo-psychrometer
Також для CPS TEMP-SEEKER TM360: Посібник користувача (20 сторінок)

1.800.561.8187

WET-SOCK METHOD

This method can be used to obtain relative humidity, dry/wet bulb and dew point mea-
surements using two standard TMX2G temperature probes and channels A and B. There
are two options for Step 4. Both options provide accurate readings and are merely a
method of user preference.
Step 1: Connect a TMX2G probe to Channel A
and another to Channel B.
Step 2: Press the RH key and then the MIN
key to display the minimum value
of the RH.
Step 3: Slip a supplied 2" cotton sock* over the
end of the TMX2G probe connected to
Channel B and wet it with distilled
water.
Step 4 (Option 1): Grip both probes approximately 18" from the ends in one hand.
Proceed to swing the probes in a circular motion through the air until the
display reading stabilizes. At that point you can stop swinging the probes.
Step 4 (Option 2): Place both probes in front of an air stream from air handler or
register. For best results the air should flow at about 9 feet per second (3 m.
second) over the wet sock ensuring that no radiant heat flows into the probe
such as direct sunlight or heat from machinery. The correct measurement is
obtained when the display stabilizes.
Step 5: Press the CLR key to erase the readings during the stabilization period and
display the current values. Pressing the RH,DB,WB or DP keys will display the
current value of the relative humidity, dry bulb, wet bulb, or dew point
temperatures.
* TMX3WB 2" cotton socks for RH measurement using wet sock method (6 pieces)
RH Range
20 to 80% RH
RH Accuracy
±3.0% RH from 20 to 80%RH and 32 to 120˚F / 0 to 50˚C
Dry & Wet Bulb Range
-50.0°F to 300˚F / -45.0°C to 150˚C
Dry & Wet Bulb
±0.4°F / ±0.2˚C from 14.0 to 158˚F / -10.0 to 70˚C; 0.6˚F / 0.3˚C
Accuracy
elsewhere in the range
DP Accuracy
+/-2.0F / +/- 1.0˚C from 32.0 to 120˚F / 0 to 50˚C and 20 to 80% RH
www.
2" cotton sock
TMX2G w/
8
.com