Install the 6 in. or 12 in. nipple into the MPG1.
Screw in the glands loosely by hand. Tighten the
gland with a 5/8 in. wrench or crescent wrench.
Install a DH500M to 1/8" NPT male adaptor in
the DUT if necessary: The 1/8 in. NPT male
adaptor can be used to connect the remaining thin
5 in. nipple to DUTs with 1/8 in NPT female ports.
If you intend to use the adaptor, wrap Teflon
thread tape around the NPT threads three times
and install it into the DUT.
Install the Device Under Test (DUT). The DUT
will be connected to the test line with the
remaining thin 5 in. nipple. This nipple can be
installed either in the 3-way tee, or in the port on
top of the MPG1. The tube can be slightly bent if
necessary. Install the nipple in the desired
location. Tighten the gland in the tee or MPG1, but
do not tighten the gland that connects to the DUT.
Plug the open test port: There is an open test port
either on the top of the MPG1, or on the 3-way tee.
Insert the stainless steel plug into a gland. Insert the
plug and gland into the open port. Do not tighten.
Prime the system, tighten fittings: Pump fluid
through the nipples by using the MPG1. See the
appropriate section of the MPG1 Operation and
Maintenance Manual for this procedure.
sure that all air is purged from the test lines.
There are 3 loose fittings: one on the DUT, one on
the RPM4 or PG7302, and the other on the plug.
Tighten each of the loose fittings when oil without
bubbles begins to flow through it.
Document 560080b
DH Instruments, Inc.
4765 East Beautiful Lane
Phoenix AZ 85044-5318
USA
The maximum operating pressure of the variable volume is
100 MPa (15 000 psi) in MPG1-100M and 200 MPa (30 000 psi)
in MPG1-200M. Do not generate pressure greater than the
maximum operating pressure.
High pressure fluids are potentially hazardous. Energy
®
stored in pressurized fluids can be released unexpectedly and with
extreme force. High pressure systems should be assembled and
operated only by personnel who have been instructed in proper
safety practices.
ALWAYS use external tubing and fittings rated for
pressures equal to or greater than the maximum pressure which
MPG1 will be used to generate.