Problem
Venting DOS cancelled.
Venting must be repeated.
Check container/siphon
DOS
Highly viscous (thick) chemicals can affect the dispenser monitoring and lead to inaccurate data. In this instance
please contact Eschmann for advice. The viscosity does not affect dosage concentration levels.

6.3. Insufficient Salt/Water Softener

Problem
Refill salt
Machine locking out
insufficient salt
Salt container empty,
programme locked out
Salt container lid not
correctly closed
MUW216743-2
Cause and remedy
Venting of the dispensing system was cancelled because an insufficient flow rate has been
identified. A dispensing hose may be kinked or the siphon blocked.
Check the dispensing hose for kinks or leaks. Position it so that it cannot be kinked.
Check the suction aperture on the siphon for blockages and remove these as necessary.
Start the venting process again.
Contact Eschmann if there are leaks in the dispensing hose or a fault with the siphon in the 5
litre container.
Little or no flow has been identified.
Check the level in the supply container and replace if necessary.
Check the suction aperture of the siphon for deposits.
Vent the dispensing system.
The dispensing hose is blocked.
Remove any kinks from the dispensing hose. Position it so it cannot become kinked.
Check the dispensing hose for leaks
Vent the dispensing system.
Contact Eschmann if there are leaks in the dispensing hose or a fault with the siphon.
Cause and remedy
Salt is running low in the water softener
Refill the reactivation salt before starting the next programme
Salt in the water softener is completely depleted and reactivation is no longer possible. The
machine is locked for further use.
Refill the reactivation salt.
The water softener cannot reactivate because there is insufficient salt. The machine is locked
for further use. Refill the reactivation salt.
The lock is lifted a couple of seconds after refilling the salt reservoir. Reactivation will occur
automatically during the next programme.
Salt residues are preventing it from closing.
Remove the residues from the refilling funnel, the lid and the seal.
Close the container properly.
19