DIALAB DTN-410 K Посібник користувача - Сторінка 17

Переглянути онлайн або завантажити pdf Посібник користувача для Вимірювальні прилади DIALAB DTN-410 K. DIALAB DTN-410 K 20 сторінок. Photometer

DIALAB DTN-410 K Посібник користувача

2.4.1 Pump volume calibration

From instrument main menu press down arrow key several times to access volume settings
and then press "1" key on built-in keyboard to start volume calibration.
MAIN MENU
PHOTOMETRICS
SETTINGS
?VOLUME
Instrument will ask for a graduate flask to perform
volume calibration:
VOLUME CALIBRATION
Insert
tip
flask with at least 5ml of
water
ASPIR
When ASPIR is pressed, several consecutive aspiration cycles will be performed and once
finished , instrument will request to write the measured volume (in uL) and internal adjust of
peristaltic turn-to-volume ratio will be calculated.
Volume calibration is recommended once a month or
when new peristaltic pump tubing is installed.
DIALAB PHOTOMETER DTN-410K USER MANUAL
MENU2
on
a
graduate
I 17 I