Dickson TH6 Посібник - Сторінка 5

Переглянути онлайн або завантажити pdf Посібник для Вимірювальні прилади Dickson TH6. Dickson TH6 20 сторінок. Temperature chart recorder

Alarm Relays (TH625)
The SPST 24V 500mA relay contacts are normally open and will close on alarm conditions when the alarms
are enabled. Relay contacts are always functional when the alarm is enabled. The relay will close only
during minimum and maximum alarm conditions.
Dip Switch Set-Up
To set-up the TH6 recorder for your specific
application, you might need to change some of the
Dip Switches. The Dip Switches are located on the
back of the unit (figure 3). Use a pen or small screw
driver to flip the switches. Remember to install
correct chart to match corresponding
Dip Switches for ranges
DICKSON TH6 Dip Switch Settings:
Up = On; Down = Off
Switches 2-4
For Model
7 Day
All Models
24 Hour
All Models
0 to +100°F
TH623/25
TH621/22
+32 to +100°F
0 to +185°F
TH623/25
0 to +50°C
All Models
0 to +85°C
TH623/25
Remember to install the correct chart for your range
Dickson - Addison, Il
Recording Time
The TH6 recorder has two different recording time
options
24 hour
7 day
NOTE: Remember to install correct chart to match
corresponding switch setting.
Rev. 10/07
#25201
Chart
2
3
4
D
U
D
D
D
D
0 to 100
D
U
0 to 200
U
D
U
U
0 to 100
www.dicksondata.com
#2 up
#2 down
Slide toggle up for on position.
1
2
Dip Switch 1,
7 & 8 are not
active on this
recorder.
Slide toggle down for off position.
Temperature Range
The TH6 will record up to 4 temperature ranges
(depended on model). Dip switch #3 & #4 allow
you to select the temperature range.
0 to +100°F
(TH623/25)
+32 to +100°F
(TH621/22)
0 to +185°F
(TH623/25)
0 to +50°C
(TH621/22/23/25)
0 to +85°C
(TH623/25)
3
4
5
6
7
8
#3 down
#4 down
#3 down
#4 down
#3 down
#4 up
#3 up
#4 down
#3 up
#4 up