Calibration mode:
First you have to open the calculator and the cover of the test bridge (next to the temperature
terminal with marking TEST) has to be removed. This invalidates the previous calibration. The
contacts of the calibration bridge, which are located in the top left corner below the protective
cover, must be briefly short-circuited (using tweezers, pointed pliers or similar).
The meter is now in the calibration mode (° sign in display).
On completion of calibration, terminate the calibration mode by short-circuiting the calibration
bridge again
(no ° sign in display). The integrator returns to normal mode automatically after the 3rd date
change.