Olympus 1X71 Посібник користувача - Сторінка 5

Переглянути онлайн або завантажити pdf Посібник користувача для Мікроскоп Olympus 1X71. Olympus 1X71 18 сторінок.

Users must always record their activity in the Log book
Transmitted light
1 . H al oge n L amp Po w e r S u pply U n i t TH 4 "O N "
2 . L i gh t Path s e le ctor on "O cu lar"
3 . K o h l e r i l l u m in a t i o n a d j u s t e d
4 . Correct phase ring in position (10x & 20x Ph1, 40x Ph2)
5 . F i l t e r c u be o n p o s it i o n # 6
Epifluorescence
Warnings:
Do not shut the unit down within 30 minutes of powering up the unit.
After shutting down the unit allow 30 minutes before re-powering up
After shutting down the unit allow 30 minutes before changing the bulb.
Failure to do so is likely to result in damage to the bulb.
1. Prior Lumen 200 module on
2. Pri or L u me n 2 0 0 inte n s ity k n ob > 0 %
3. Light Path selector lever on "Ocular"
4. Corre ct f i l te r cu be in pos ition
5. Fluorescence shutter open
Image Capture
1. Start Camera controller (press until LED turns green)
2. Compu te r on (L o gi n : J an e / Pas s word : J a* e )
3. H Clmage s of tware ope n
4. L i gh t Path s e le ctor le v e r on Came ra
5. Correct transmitte d ligh t/e pifluorescence set -up
6. " L i v e " m o d e
7. Adjust exposure time accordingly. Make use of the Histogram and the Saturatio n
options
8. " A b o r t "
9. " C a p t u r e l "
1 0 .
S a v e a s i n M y D o c u m e n t s > U s e r N a m e _ U n i t > F i l e N a m e . t i f
1 1 .
Shut-down: exit HClmage, log out windows session, camera on stand -by (press
until LED turns orange)

Quick user guides

5