Olympus BH-2 BHT Посібник - Сторінка 16
Переглянути онлайн або завантажити pdf Посібник для Мікроскоп Olympus BH-2 BHT. Olympus BH-2 BHT 24 сторінки. Sliding focus block
Також для Olympus BH-2 BHT: Повний демонтаж, очищення та повторний монтаж (23 сторінок)
Figure 18 – Lamp House Plugged into BHT/BHTU Electrical Base (100/115V Version)
A repair kit for the Olympus lamp house is available from J.C. Ritchey Company, LLC (see
Figure
19). This kit, which
includes a replacement for the back cover and a new G4 lamp socket, may be used to overhaul either the LS20H or the
LS20H‐M2 lamp houses. The replacement back is provided since the original can be damaged during disassembly. The
installation of this repair kit requires soldering, but defective lamp houses may be sent to J.C. Ritchey Company for in‐
house repair/exchange.
Figure 19 – BHT/BHTU Lamp House Repair Kit from J.C. Ritchey Company (photo courtesy James Ritchey)
Olympus BH‐2 (BHT/BHTU) Electronics
Revision 2
Page 16 of 24