Olympus BHT-111 Інструкція з експлуатації - Сторінка 17

Переглянути онлайн або завантажити pdf Інструкція з експлуатації для Мікроскоп Olympus BHT-111. Olympus BHT-111 22 сторінки. Bht series

Use the field of view eyepieces for focusing on the film plane. Each field of vieweyepie
has a focusing front lens and a reticle with 4 frames, each frame indicating the area cover
by a specif ic power N F K photo eyepeice. (Fig. 22!-.
The number at each frame indicates the
magnification of the photo eyepiece. The
image in the field of view eyepiece and the
image on the film plane are in focus at the
same time. Several type field of view eye-
pieces are available, according to the film
size employed.
Field of view eyepiece
35WH Kl OX
PWH Kl OX
4X5WH K 1OX
MHWH Kl OX
Attachment camera
3 5mm B a ck
3T+" x 4Y+"
P ol a ro i d
B ac k
4" x5 "
S h e et F i lm or
P ol a ro i d F i lm
H ol d er
1 6 m m B o l e x
camera
1 2 0 R o l l F i l m
H o l d e r
1) Select the field of view eyepiece matching the camera back in use and insert it into th
right eyepiece tube of the trinocular tube, aligning locating groove and locating pin.
2) While looking through the field of view eyepiece, rotate the eyepiece front lens in scre
mount to focus on the double cross lines in the field. For sharp focusing with objectiv
4X or lower, the focusing magnifier FT is recommended.
3) Bring the specimen detail to be photographed within the frame corresponding to th
power of the NFK eyepiece in use and focus on the specimen with the microscope fin
adjustment knobs. Make sure the light path selector knob on the observation tube
either on the white (V) or yellowgreen (CV) band.
4) lt is recommended to tighten the tension adjustment ring considerably to prevent th
stage from dropping during long exposures.
Fig.22
1 5