c) The image is not
focused.
When adjusting the Stage height, you
forgot to reattach the upper stopper
screw.
Reattach the upper stopper screw.
d) Coarse adjustment will
not go all the way up.
You are trying to raise the Stage with
the coarse adjustment
knob with the
pre-focusing lever engaged.
Unlack the pre-focusing lever.
18
e) Coarse adjustment will
not go all the way down,
The condenser holder is too low.
Raise the condenser holder
f) The objective
makes
contact with the speci-
men before focus
is
obtained.
The specimen is mounted upsidedown
Mount the specimen correctly.
14
;
Adjust the interpupillary distance.
4. Observation
Tube
_'
a) Field of view of one eye
does not match that of
the other.
The interpupillary distance is incorrect.
Incorrect diopter adjustment
Adjust the diopter.
14
Different eyepieces are used on the left
and right.
Change one eyepiece to match the other
so that both sides are the same type.
The Optical axes are not parallel.
Upon looking into the eyepieces,
try
looking at the Overall field before con-
centrating on the specimen range. You
may also find it helpful to look up and
into the distance for a moment before
looking back into the microscope.
3
><
",n
:
~>
The Stage is not properly mounted.
5. Stage
a) The image shifts when
you tauch the Stage.
Clamp the Stage.
b) Specimen
Stops mid-
way
on the
X axis
traverse.
eR-~e specimen is not correctly position-
Place the specimen correctly.
12
c) The X axis and Y axis
knobs are too tight, or
too loose.
Adjust the tension.
Is X or Y axis tension too high or too
low?
12