Olympus BX43 Інструкція з експлуатації - Сторінка 28

Переглянути онлайн або завантажити pdf Інструкція з експлуатації для Мікроскоп Olympus BX43. Olympus BX43 48 сторінок. System microscope

Olympus BX43 Інструкція з експлуатації
1

5-5 Condenser

For how to adjust the centering of the condenser, see page 12.
1

Compatibility of Objectives and Condensers

Objective
Magnification
1.25X
2X
4X
10-60X
100X
* When using the U-SC3 swing-out condenser together with a 1.25X to 4X objective, fully open the condenser aperture
iris diaphragm and use the field iris diaphragm in the base as aperture diaphragm. With the 1.25X to 2X objectives, the
surroundings of the field of view may become dark.
} To obtain better illumination in camera recording using a 1.25X to 4X objective, use of the U-ULC-2 is recommended.
**When the U-LC is used with an objective with an NA of 1.0 or more, the NA of the lighting becomes insufficient.
} When the U-LC is used, it is required to attach the auxiliary lens (page 34).
24
2
Fig. 39
Low
magnification
U-LC
Applicable
(FN 22)**
6

Using Eyepieces Incorporating a Micrometer

} When the eyepieces in use incorporate a micrometer, the accuracy of
the left-right focusing adjustment (diopter adjustment) can be improved
further.
1. Looking into the right eyepiece with your right eye, turn the top of the
eyepiece @ so that the micrometer in the field of view looks sharpest
(Fig. 39).
2. Looking into the right eyepiece, rotate the coarse and fine adjustment
knobs to bring the specimen and micrometer (cross lines, etc.) in focus.
3. Looking into the left eyepiece with your left eye, rotate the diopter
adjustment ring 2 to bring the specimen into focus.
Condenser
Abbe
Achromat/aplanat
U-AC2
U-AAC
Applicable
(FN 22)
Applicable
Applicable
(FN 26.5)
(FN 26.5)
Ultralow
Swing-out
magnification
U-SC3
U-ULC-2
Applicable
(FN 22)
Applicable
Applicable (FN 26.5)
(FN 26.5)
by moving top lens
out of the light path*
(FN 26.5)
Engage top lens in