Olympus SZX16 Інструкція з експлуатації - Сторінка 6

Переглянути онлайн або завантажити pdf Інструкція з експлуатації для Мікроскоп Olympus SZX16. Olympus SZX16 36 сторінок. Research high-class stereo microscope
Також для Olympus SZX16: Огляд (24 сторінок), Інструкція з експлуатації (28 сторінок), Інструкція з експлуатації (40 сторінок)

Olympus SZX16 Інструкція з експлуатації
1
Getting Ready
2
Fig. 2
²
³
@
Fig. 3
1. A microscope is a precision instrument. Handle it with care and avoid
subjecting it to sudden or severe impact.
2. Do not use the microscope where it is subjected to direct sunlight, high
temperature and humidity, or vibration. (For the operating environment,
see Chapter 6, "SPECIFICATIONS" on page 16.)
3. When moving the microscope, remove the specimen in advance. Then,
hold the front part of the base with one hand and hold the rear part of the
base with the other hand to avoid tilting the microscope.
Also remove any module or attachment from the microscope to minimize
the weight before moving.
4. Observe the following cautions when operating the coarse focus
adjustment knob or the zooming knob.
Operation
Manipulated Controls
Focusing
Coarse/fine focus
adjustment knobs @
(Fig. 3)
Zooming
Zooming knob ²
(Fig. 3)
5. Only one intermediate attachment can basically be used, but two
attachments may also be used though the peripheral part of the image
may be obscured.
For the coaxial vertical illuminator (SZX2-ILLC16) and reflected fluorescent
light illuminator (SZX2-RFA16), only either can be attached because they
must always be attached at the lowest position.
}The SZX2-ILLC16 is not considered to be an attachment. Therefore, the
SZX2-LBS can be mounted above it without any restriction.
Caution
1. If the knob hits the upper or lower
limiting mechanism violently or
it is rotated after it hits a limiting
mechanism, the internal
mechanism may be damaged.
2. If the knobs on the left and right
are rotated in opposite directions,
the internal mechanism will be
damaged. (The rotation tension
of the coarse focus adjustment
knob should be adjusted using
the rotation tension adjustment
ring ³ on the knob.)
1. If the knob hits the upper or lower
limiting mechanism violently or
it is rotated after it hits a limiting
mechanism, the internal
mechanism may be damaged.
2. If the knobs on the left and right
are rotated in opposite directions,
the internal mechanism will be
damaged.