Olympus U-AW Інструкція з експлуатації - Сторінка 12

Переглянути онлайн або завантажити pdf Інструкція з експлуатації для Мікроскоп Olympus U-AW. Olympus U-AW 16 сторінок. Motorized attenuator wheel

Olympus U-AW Інструкція з експлуатації
2

Installation in a Transmitted Light Illumination System

When using BX63
1
Loosen the U-LH100ADP clamping screws a and b , and separate the U-LH100ADP to 3 units: [Front], [Middle] and
[Rear]. (The [Middle] unit is not used.)
2
Loosen the lamp housing clamping screw c of the microscope frame using the Allen screwdriver, fit the mount
dovetail d of the U-LH100ADP [Front] and tighten the clamping screw c .
3
Attach the U-AW e to the U-LH100ADP [Front] and tighten the clamping screws a .
4
Loosen the clamping screws f (4 pcs.) of the U-AW, fit the U-LH100ADP [Rear] g by aligning the positioning pin,
and tighten the clamping screws f (upper 2 pcs. only).
5
Attach the lamp housing h to the U-LH100ADP [Rear] g in the same way as above.
6
Insert one end of the interface cable into the connector i and insert the other end into the BX63F microscope frame
or control box (see "SYSTEM CHART").
When U-AW is not mounted properly, U-AW may not be recognized in some cases. If it is not recognized,
CAUTION
mount U-AW properly.
8
e
a
d
c
i
h
g
f
Positioning pin
This adapter (the part without
lens) is not used.
b