Attaching the Lamp Housing (Fig. 8)
1. Loosen the two lamp housing clamping screws @ of the UV248 light
source box.
2. Insert the mount seat ² of the lamp housing into the mount and tighten
the clamping screws @.
Attach the lamp housing and take care not to tilt it. Reserve enough
space around the lamp housing to ensure ventilation.
Resetting the Burner Hour Counter
Set the hour counter by referring to the instruction manual provided with
the power supply.
4
Attaching the UV Light Guide
Do not bend the UV light guide into a radius below 100 mm to
prevent it from breaking. Also do not place a heavy object on the
light guide or subject it to strong force.
1. Remove the protective rubber caps from the two ends of the light guide.
Handle the light guide carefully so as not to leave stain or damage on
the end surfaces.
2. Insert the input connector @ into the UV light guide mount ² on the
UV248 light source box by aligning the positioning marks, and tighten
the clamping ring ³.
3. Loosen the clamping knob | and insert the output connector ƒ all
the way. Clamp the connection so that the clamping knob | is aligned
with the positioning reference plane ....
5
Attaching the MApo100X Objective for UV Observation
Attach the objective in the same way as a UIS2 (UIS) objective.
To prevent unexpected irradiation of UV rays, be sure to close the
idle objective positions of the revolving nosepiece by attaching the
caps provided with the revolving nosepiece.
(Figs. 9 & 10)