IMPORTANT

SAFETY PRECAUTIONS
Fig. 1

Safety Symbols

The following symbols are found on this system. Study the meaning of the symbols and always use the equipment in
the safest possible manner.
Symbol
1
} This manual pertains only to the information related to the scanning stage
with ultrasonic and associated modules.
Before using this instrument together with the BX63 microscope and
associated modules, make sure that you have carefully read their manuals
and understand how the system should be operated together.
CAUTION
1. Install the microscope on a level (inclination less than ±2°) surface on a
sturdy table or bench.
2
2. To prevent obstruction of the air vents 2 on both sides of the control box
@ for BX3-SSU, always install it longitudinally and make sure to leave at
least 10 cm of free spaces around them.
Also, to prevent malfunction or electric shock, do not allow a wire or thin
object to enter the air vents.
CAUTION
1
3. Always use the power cord provided by Olympus. If no power cord is
provided, please select the proper power cord by referring to the section
"PROPER SELECTION OF THE POWER SUPPLY CORD" at the end of this
instruction manual. If the proper power cord is not used, product safety
performance cannot be warranted.
4. Always ensure that the grounding terminal of the instrument and that of
the wall outlet are properly connected. If the equipment is not grounded,
Olympus can no longer warrant the electrical safety performance of the
equipment.
5. The power cord can be used to shut down the power supply in case of
emergency. To make this possible, place the power cord plug in an easily
accessible location.
Indicates a non-specific general hazard. Follow the description given after this symbol or in
instruction manual.
Indicates that the main switch is ON.
Indicates that the main switch is OFF.
Indicates a potential fire hazard; when replacing fuses, be sure replacement fuses are of the
specified rating.
To ensure the performance, we recommend that you have your
Olympus representative assemble and adjust the scanning
stage with ultrasonic.
When installing the BX3-CBH and the control box for BX3-SSU
side by side, always install the BX3-CBH on the left to enable
effective heat radiation.
Explanation
(Fig. 1)